We are proud to introduce the ULTRA GI™ Cycle - a newly added cycle to the STERRAD™ 100NX Sterilization System with ALLClear™ Technology.

This milestone allows healthcare providers to use the trusted STERRAD™ 100NX System to terminally sterilize hydrogen peroxide-compatible duodenoscopes, helping enhance infection prevention in gastrointestinal (GI) procedures. 

 

✅ Sterilize Duodenoscopes

  • STERRAD™ 100NX System can now* sterilize compatible duodenoscopes with the following specifications: Lumens ≥1 mm and ≤1500 mm in length
  • Lumens ≥2 mm and ≤1630 mm in length
  • Instruments with up to 4 channels 

The ULTRA GI™ Cycle expands the STERRAD™ System’s capabilities into GI endoscopy—providing a validated terminal sterilization alternative to high-level disinfection.

This cycle can sterilize hydrogen peroxide compatible duodenoscopes in the new APTIMAX™ Instrument Tray for ULTRA GI™ Cycle along with validated sterilization wrap. 
The STERRAD VELOCITY™ BI can now be placed into the new single-use ULTRA GI™ PCD Vial for periodic testing and frequent monitoring of the ULTRA GI™Cycle.

STERRAD™ ULTRA GI™ Cycle is setting a new standard in endoscope reprocessing!

  • NEW! Terminal Sterilization for GI Endoscopes
  • Fast & Reliable Sterilization: 68-Minute Cycle
  • Upgradeable Future-Ready Platform STERRAD™ 100NX with ALLClear™ Technology offers upgradeable features, protecting your investment as standards evolve.
  • Innovation with Validated Accessories
  • Easy Monitoring for Peace of Mind
STERRAD™ ULTRA GI™ Cycle Unveiled
53886
PDF Icon PDF document

How can I have ULTRA GI™ Cycle in my sterilization facility?

Upgrade your existing STERRAD™ 100NX System with ALLClear™ Technology

OR

Purchase a new STERRAD™ 100NX System with ALLClear™ Technology with the cycle pre-installed

Why ULTRA GI™ Matters: A Breakthrough in Infection Prevention

Even with current high-level disinfection protocols, endoscope-related contamination rates remain high-up to 16.1%. Complex designs, manual processes, and human error contribute to infection risk.

  • Key Fact: Infections linked to duodenoscope use continue to affect high-risk populations, especially in ERCP procedures.

Why Duodenoscopes Demand More

  • Intricate designs with elevator mechanisms make cleaning difficult.
  • Outbreaks from improperly reprocessed duodenoscopes have triggered FDA mandates and Congressional investigations.
  • Studies show persistent contamination even after proper HLD.

The ULTRA GI™ Cycle targets these risks with a sterilization method validated for duodenoscopes.

Environmental & Operational Impact

While single-use scopes eliminate cross-contamination risk, they:

  • Generate 40% more medical waste
  • Have a 2.3x higher carbon footprint
  • Raise cost and supply chain concerns

ASP’s ULTRA GI™ Cycle offers a sustainable solution: enabling reuse of scopes with effective, low-temp sterilization.

Sterilization vs. Disinfection

Disinfection reduces most pathogens but does not eliminate spores. Sterilization, as enabled by the ULTRA GI™ Cycle, destroys all microbial life, including highly resistant spores.

MethodMicrobial ReductionEliminates Spores?
Cleaning + HLD6 log₁₀❌ No
Sterilization12 log₁₀✅ Yes

High-Risk Patients Need a Higher Standard 

  • ERCP patients are often immunocompromised-transplant recipients, cancer survivors, elderly.
  • They deserve the highest margin of safety.
  • Clinical Findings: Up to 5.5% of patients had infections within 30 days post-ERCP.
  • 97% of transplant patients undergoing ERCP required therapeutic intervention.

View White Paper for the full details on How ULTRA GI™ Reduces GI Endoscope Infections
53891
PDF Icon PDF document

Data Sheet: Be the Expert, Lead the Market!

✅ Game-Changer for Duodenoscope Sterilization 

  • Terminal sterilization of compatible duodenoscopes - up to 4 channels
  • Delivers Sterility Assurance Level (SAL) 10⁻⁶ with a validated double-kill cycle 

Powered by ALLClear™ Technology 

  • Integrated moisture management & load warming - built into the cycle 

✅ Engineered for Efficiency 

  • 68-minute cycle time with optimized lumen claims (≥1mm × ≤1500mm; ≥2mm × ≤1630mm)
  • Up to 2 duodenoscopes per cycle

✅ Reliable Sterilant Delivery System

  • 6 H₂O₂ injections per cycle in a closed cassette system - automatic expiration detection

✅ Total Compliance Confidence

  • Designed to align with globally recognized guidelines for duodenoscope reprocessing

✅ Specialized Tools for Best Results

  • ULTRA GI™ PCD Vial for sterility assurance
  • APTIMAX™ Instrument Tray - tailored for duodenoscopes

 

*CE mark pending

ULTRA GI™ Cycle Data Sheet
53906
PDF Icon PDF document

Stay Connected

Start the conversation on increased safety in your sterilization department now!

Disclaimer

The content of this website is aimed exclusively at healthcare professionals (HCPs) and not at the general public. Please confirm by clicking below that you are a healthcare professional, duly authorized to practice in your field. ASP EMEA declines all responsibility in case of unauthorized consultation.